One catheter a month. One touch to control.

A balloon-free indwelling urethral catheter for female patients with voiding dysfunction. A quatrefoil tip holds it in place, and an integrated ball-bead valve lets the patient or carer drain the bladder by hand, then close it again.

IndicationFor female patients with voiding dysfunction. Not indicated for male patients. Registered in China only.
Disposable Sterile Controlled Catheter: silicone shaft with quatrefoil tip, sealing ring, baffle and ball-bead valve
One monthOne catheter per month of use
One touchValve opened and closed by hand
No balloonQuatrefoil tip holds the catheter in place
11 modelsW35–W55 lengths · Fr 12–20
How it works

Retention without a balloon. Drainage on demand.

Conventional indwelling catheters drain continuously into a bag and are held by an inflated balloon. The Silicon Life catheter takes the opposite approach: it lowers outflow resistance only when the patient chooses to void. With the valve open, the bladder empties through the catheter under the pressure difference and gravity; with the valve closed, the catheter stays sealed between voids.

Valve closed · ball bead seated, catheter sealed between voids
INSIDE · BLADDER NECK AND URETHRAOUTSIDE · URETHRAL MEATUSØ 4.0–6.7 mmSEALEDBULGED SECTION12345TIP END VIEW
Fig. 1 · Side elevation, schematic · not to scalePush rod (6) not shown; it is removed after placement
  1. 1
    Quatrefoil tip
    Four flexible lobes retain the catheter at the bladder neck.
  2. 2
    Shaft
    Silicone rubber. Effective length 35, 45 or 55 mm; Fr 12, 16 or 20.
  3. 3
    Sealing ring
    Standard, or reinforced on Max models.
  4. 4
    Ball-bead valve
    PTFE ball bead in a bulged section of the shaft.
  5. 5
    Baffle
    Rests outside the urethral meatus and prevents migration.
  6. 6
    Push rod
    Stainless steel (06Cr19Ni10) insertion aid, inserted from the outer end and removed after placement.
01
Balloon-free quatrefoil tip

A four-lobed tip sits at the bladder neck and keeps the catheter in place without an inflatable balloon.

02
Ball-bead valve

A PTFE ball bead seals the lumen. Moved into the bulged section of the shaft, it opens the outflow; moved back, it closes it.

03
One-touch control

The valve is operated by squeezing the outer end of the catheter. No bag, clamp or inflation syringe is needed between voids.

The Disposable Sterile Controlled Catheter held between two fingers, showing the ball-bead valve end, the transparent silicone shaft and the quatrefoil tip
The actual device. Still from the company’s operation video.
EN · CAPTIONS9:16
Introduction
00:0000:50
Download MP4
Instructions for use

Step by step

The operation video follows the steps below. Select a step to jump to that chapter. Always follow the printed Instructions for Use.

  • The catheter is placed and replaced by a healthcare professional at a medical institution.
  • It remains in place for up to 30 days; the patient then returns for a new catheter.
  • The patient or carer must be able to operate the valve by hand.
Specifications

11 models

Three effective lengths and three French sizes. Max models add a reinforced sealing ring.

Length
Size
SHOWING 11 OF 11
Controlled catheter models and dimensions
ModelEffective lengthSizeShaft diameterSealing ring
W35-Fr1235 mmFr 124.0 mmStandard
W45-Fr1245 mmFr 124.0 mmStandard
W35-Fr1635 mmFr 165.3 mmStandard
W45-Fr1645 mmFr 165.3 mmStandard
W55-Fr1655 mmFr 165.3 mmStandard
W35-Fr2035 mmFr 206.7 mmStandard
W35-Fr20MaxMAX35 mmFr 206.7 mmReinforced
W45-Fr2045 mmFr 206.7 mmStandard
W45-Fr20MaxMAX45 mmFr 206.7 mmReinforced
W55-Fr2055 mmFr 206.7 mmStandard
W55-Fr20MaxMAX55 mmFr 206.7 mmReinforced

W35, W45 and W55 give the effective shaft length in millimetres. Shaft diameter is the cylindrical shaft diameter for each French size. Select a longer length if the patient feels urethral distension, and a larger diameter if leakage occurs.

Materials and sterilisation

As stated on the registration certificate.

Shaft (baffle, sealing ring, quatrefoil tip)
Silicone rubber
Ball bead
Polytetrafluoroethylene (PTFE)
Push rod
Stainless steel 06Cr19Ni10
Sterilisation
Ethylene oxide; supplied sterile
Indications

Indicated for female patients with voiding dysfunction

Indicated for

Female patients with voiding dysfunction, including those for whom indwelling or intermittent catheterisation is being considered.

Contraindications
  • Known allergy to any product material.
  • Severe urethritis.
  • Male patients — the device is not indicated for men.
Precautions
  • If the patient feels urethral distension or has difficulty voiding, select a model with a longer effective length.
  • If leakage occurs, select a model with a larger shaft diameter.
  • Read the Instructions for Use before placement.
Patient groups
  • Neurogenic bladder after spinal cord injury
  • Voiding difficulty after radical pelvic surgery
  • Bladder emptying disorders in multiple sclerosis, Parkinson’s disease or diabetic neuropathy
  • Voiding difficulty after stroke

Patient groups described by the company. Selection criteria include adequate bladder capacity and the hand function needed to operate the valve.

Registration

Medical device registration certificate

Authority
Hubei Medical Products Administration (NMPA framework)
Country
People’s Republic of China
Device class
Class II
Registrant
Wuhan Guiji Shengming Medical Technology Co., Ltd.
Registration no.
鄂械注准20252145807
Scope
Female patients with voiding dysfunction
Approved
17 October 2025
Valid until
16 October 2030

This device is registered for the Chinese market only. It is not registered or approved for sale in any other market.

Clinical study

Study summary

A single-centre clinical study with a small sample and no control group. Results should be read with this in mind.

Request the full study report→
Design
Interventional, single-centre, single-arm
Enrolled / completed
23 / 19 participants
Duration
One catheter per participant for 30 days
Population

Women unable to void voluntarily (detrusor underactivity, detrusor–sphincter dyssynergia, central nervous system injury, stress incontinence or post-stroke voiding dysfunction)

Primary endpoint

Convenience and comfort of placement, voiding and indwelling (combined score ≥ 4)

Secondary endpoint

Serious adverse events and adverse events

Result

In the per-protocol set, 18 of 19 participants (94.7%; 95% CI 74.0–99.9%) reached the convenience and comfort threshold, against a prespecified target of 70% (p = 0.0186).

Safety

No adverse events occurred among the 19 participants who completed the study. Four participants withdrew after the catheter dislodged or was removed because of discomfort; all four events were moderate and non-serious and resolved without sequelae. The investigators attributed them to a mismatch between patient anatomy and the two catheter sizes used in the study.

Limitations

Small, single-centre study without a control group. Only two models (6.5 mm shaft; 35 and 45 mm length) were studied, which led the investigators to recommend a wider size range.

Site: Tongji Hospital, Tongji Medical College, Huazhong University of Science and TechnologyPeriod: January – June 2025Sponsor: Wuhan Guiji Shengming Medical Technology Co., Ltd.
Downloads

Documents and media

  • MP4
    Operation video
    English captions · MP4 · 0:50
    Download↓
  • JPG
    Registration certificate
    Chinese · JPG
    Download↓
  • DOC
    Instructions for Use
    For healthcare professionals · on request
    Request access
  • DOC
    Clinical study report
    For healthcare professionals · on request
    Request access

The Instructions for Use and the clinical study report are provided to healthcare professionals on request.